TECHNICAL CHARACTERISTICS
- Operational mode : Automatic (synchro) with memory of 10s per lead, Manual (real time)
- Leads (channel) : 12
- Printout format : 1, 2, 2+R, 3 channels
- Printer : Digital thermal printing ( high resolution) 50mm
- Paper width : 60mm
- Paper type : Thermal Black Rapid Z- folded paper - width 60mm X 20 meters
- Paper speed : 25 – 50mm/s ±5%
- Display : LCD 32 x 120 pixel
- Sensitivity : 5 – 10 – 20 - auto (mm/mV) ±5%
- Time constant : > 3,3s
- Copy function : Last ECG recording ( 20+ ECG optional)
- Signal calibration : Automatic
- Electrodes check : Automatic ( impedance )
- Trace centring : Automatic ( always active )
- Frequency range : 0,05 – 150Hz ( -3dB) without filter
- Filters : 50/60 Hz modified Notch, ADF (always active)
- Input Impedance : >100Mohm
- Pacemaker detection : Automatic ( IEC rules)
- Acquisition : 11 bit
- CMRR : 90dB
- Keyboard : Alphanumeric
- Patient data : Yes ( print on paper)
- Time and date : Yes ( print on paper)
- Parameter calculation : Yes ( optional)
- Rhythm analysis : Yes (optional)
- E.C.G. Analysis : Yes. Adult (optional)
- Wireless connection : IrDA (Infrared ray)
- Defibrillation protection : Inside the device
- Heart Rate : Yes ( On display and print on Paper)
- Power supply : Universal (115 o 230 Volt ) and Accu Rechargeable
- Accumulator autonomy : 3 hours (1channel mode)
- Dimensions and Weight : 250x174x52mm - 1kg with accumulator
- Security and Conformity : Class II, CF type. Standards : 89/336/EEC, 93/68/EEC, MDD 93/42/EEC - CE 0470
|
|